PTCA球囊扩张导管关键技术要求英文翻译.doc
《PTCA球囊扩张导管关键技术要求英文翻译.doc》由会员分享,可在线阅读,更多相关《PTCA球囊扩张导管关键技术要求英文翻译.doc(10页珍藏版)》请在咨信网上搜索。
1、Technical requirements for medical device product registration number:Disposable Sterile PTCA balloon catheter1、Model / Specifications and divide instructions1、1 Model PTCA-XX-XXBalloon Working Length mmBalloon nominal Diameter mmProduct Name CodeModel mark example:PTCA2.010 represents balloon nomin
2、al outer diameter is 2.0mm and working length is 10mm of disposable Sterile PTCA balloon catheter1、2 Specifications PTCA balloon catheter specification listSpecificationsODLengthThe effective length of the catheterMax guide wire diameterNominal pressureRBPDK-PTCA-1.5-101.50mm10mm142cm0.014in8ATM16AT
3、MDK-PTCA-1.5-201.50mm20mm142cm0.014in8ATM16ATMDK-PTCA-1.5-301.50mm30mm142cm0.014in8ATM16ATMDK-PTCA-1.75-101.75mm10mm142cm0.014in8ATM16ATMDK-PTCA-1.75-201.75mm20mm142cm0.014in8ATM16ATMDK-PTCA-1.75-301.75mm30mm142cm0.014in8ATM16ATMDK-PTCA-2.0-102.00mm10mm142cm0.014in8ATM16ATMDK-PTCA-2.0-202.00mm20mm14
4、2cm0.014in8ATM16ATMDK-PTCA-2.0-302.00mm30mm142cm0.014in8ATM16ATMDK-PTCA-2.5-102.50mm10mm142cm0.014in8ATM16ATMDK-PTCA-2.5-202.50mm20mm142cm0.014in8ATM16ATMDK-PTCA-2.5-302.50mm30mm142cm0.014in8ATM16ATMDK-PTCA-2.75-102.75mm10mm142cm0.014in8ATM16ATMDK-PTCA-2.75-202.75mm20mm142cm0.014in8ATM16ATMDK-PTCA-2
5、.75-302.75mm30mm142cm0.014in8ATM16ATMDK-PTCA-3.0-103.00mm10mm142cm0.014in8ATM16ATMDK-PTCA-3.0-203.00mm20mm142cm0.014in8ATM16ATMDK-PTCA-3.0-303.00mm30mm142cm0.014in8ATM16ATMDK-PTCA-3.5-103.50mm10mm142cm0.014in6ATM16ATMDK-PTCA-3.5-203.50mm20mm142cm0.014in6ATM16ATMDK-PTCA-3.5-303.50mm30mm142cm0.014in6A
6、TM16ATMDK-PTCA-4.0-104.00mm10mm142cm0.014in6ATM16ATMDK-PTCA-4.0-204.00mm20mm142cm0.014in6ATM16ATMDK-PTCA-4.0-304.00mm30mm142cm0.014in6ATM16ATM1、3 Divide instructionsDisposable Sterile PTCA balloon catheter consists of balloon,connecting catheter,tip,support catheter and Handle. According to differen
7、t balloon diameter to divide the balloon into different specifications and sizes. According to medical device classification rules 6877,the balloon catheter belonging to Class interventional equipment intravascular catheter.2、Performance2、1 AppearanceAs with normal vision or corrected vision at a ma
8、gnification of 2.5 times the conditions of inspection,the effective length of the outer catheter should be clean and free of impurities. There should be no processing defects and surface defects on the surface of catheter effective length and end.The inner and outer surface of the distal end compone
9、nts of the catheter should have silicone oil (polydimethylsiloxane) lubricant coating. With normal vision or corrected vision inspection,the outer surface of the catheter should not see the convergence of lubricant droplets.Note:The inner and outer surface of the distal end components of the cathete
10、r is the out surface from the tip (figure 1) to the transition weld (figure 7),and the inner surface from guide wire lumen (from the tip (Figure 1 below) to quickly switch ports (figure 5).2、2 Size requirements2.2.1 Balloon catheter size requirements reference table1,specification list reference tab
11、le 2.1、Tip 2、Balloon 3、Radiopaque ring 4、Out catheter 5、Quickly switch ports 6、Guide wire 7、Transition weld 8、Hypo Tube(FEP Coated) 9、Bending resistant tube 10、Handle 11 Balloon effective length(L) 12、catheter effective length Pic 1. PTCA balloon catheter schematicPic 2. Tip schematicTable 1 disposa
12、ble Sterile PTCA balloon catheter size and tolerance requirementsNameUnitStandardToleranceRemarkBalloon diameter under nominal pressuremmD0.125Balloon effective lengthmmL10%Assembly integrated tolerances10%,Reference pic 1.Radiopacity Length(Platinum ring spacing)mm(L-1)mm(L is balloon effective len
13、gth)10%Adapted to the maximum diameter of the guide wireInchNANA0.014in Guide wire pass through guide wire lumen smoothly(Wear into the tip,piercing out from fast switching port 5)Max catheter OD in componentsmm1.00.1Proximal welding pointCatheter effective lengthcm1422Reference pic 1.Tip sizemm20.5
14、Reference pic 2. outside of crater2.2.2 Tapered connectorConnector of balloon catheter must be tapered connector. Size should be consistent with GB / T 1962.1- requirements.2.3 Physical properties2.3.1 Breaking forceBreaking strength of each test section shall comply with the requirements of Table 3
15、.Table 3 Catheter test section breaking forceMin OD of catheter test section /mmMinimum breaking force /N0.550.550.750.751.151.151.851.851.38351015Note:YY0285.1- did not make provisions for the breaking force which catheter OD is less than 0.55mm.2.3.2 Leakagea) There should no liquid leakage at cat
16、heter hub,connection assembly part and other parts of catheter. There should be no leak or signs of damage for catheter or balloon. If prominent or burst,this requirement adapt before hydrated and after hydrated.b)In the continuous process of the suction,air should not enter into the assembly part o
17、f catheter. This requirement adapt before hydrated and after hydrated.2.3.3 Ray detectabilityRadiopaque ring should be visible under proven imaging technology which in the clinical use.2.3.4 Tip configurationIn order to reduce damage to blood vessels during the use process,the end of tip should be s
18、mooth and have a taper or there is a certain process or via similar finishing.2.3.5 Particulate pollution Balloon catheter production should under the min particulate pollution conditions,inner and out surface of catheter should be clean and pollution index shall comply with the requirements of GB 8
19、368- in Chapter 6.1.2.3.6 Non-hydratable Changes of effective before and after hydration should not exceed 4mm or 1% higher than before hydration(Take min);and max OD changes before and after hydration should not exceed 10%of max OD.2.4 Chemical properties2.4.1 Corrosion resistanceShall comply with
20、the requirements 4.4 of YY 0285.1-.2.4.2 Reducing substance(Readily oxidizable substance)The volume difference between test liquid and blank liquid consume potassium permanganate solution c (KmnO4) = 0.002mol / l shall not exceed 2.0ml.2.4.3 cThe total content of barium,chromium,copper,lead,tin in t
21、he test liquid should not exceed 1g / ml,Cadmium content should not exceed 0.1g / ml.2.4.4 PHThe difference between test pH value and the control of pH value should be less than 1.5.2.4.5 Residual ethylene oxideResidual ethylene oxide of balloon catheter should not more than 10ug/g.2.4.6 Balloon cat
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- PTCA 扩张 导管 关键技术 要求 英文翻译
1、咨信平台为文档C2C交易模式,即用户上传的文档直接被用户下载,收益归上传人(含作者)所有;本站仅是提供信息存储空间和展示预览,仅对用户上传内容的表现方式做保护处理,对上载内容不做任何修改或编辑。所展示的作品文档包括内容和图片全部来源于网络用户和作者上传投稿,我们不确定上传用户享有完全著作权,根据《信息网络传播权保护条例》,如果侵犯了您的版权、权益或隐私,请联系我们,核实后会尽快下架及时删除,并可随时和客服了解处理情况,尊重保护知识产权我们共同努力。
2、文档的总页数、文档格式和文档大小以系统显示为准(内容中显示的页数不一定正确),网站客服只以系统显示的页数、文件格式、文档大小作为仲裁依据,个别因单元格分列造成显示页码不一将协商解决,平台无法对文档的真实性、完整性、权威性、准确性、专业性及其观点立场做任何保证或承诺,下载前须认真查看,确认无误后再购买,务必慎重购买;若有违法违纪将进行移交司法处理,若涉侵权平台将进行基本处罚并下架。
3、本站所有内容均由用户上传,付费前请自行鉴别,如您付费,意味着您已接受本站规则且自行承担风险,本站不进行额外附加服务,虚拟产品一经售出概不退款(未进行购买下载可退充值款),文档一经付费(服务费)、不意味着购买了该文档的版权,仅供个人/单位学习、研究之用,不得用于商业用途,未经授权,严禁复制、发行、汇编、翻译或者网络传播等,侵权必究。
4、如你看到网页展示的文档有www.zixin.com.cn水印,是因预览和防盗链等技术需要对页面进行转换压缩成图而已,我们并不对上传的文档进行任何编辑或修改,文档下载后都不会有水印标识(原文档上传前个别存留的除外),下载后原文更清晰;试题试卷类文档,如果标题没有明确说明有答案则都视为没有答案,请知晓;PPT和DOC文档可被视为“模板”,允许上传人保留章节、目录结构的情况下删减部份的内容;PDF文档不管是原文档转换或图片扫描而得,本站不作要求视为允许,下载前自行私信或留言给上传者【快乐****生活】。
5、本文档所展示的图片、画像、字体、音乐的版权可能需版权方额外授权,请谨慎使用;网站提供的党政主题相关内容(国旗、国徽、党徽--等)目的在于配合国家政策宣传,仅限个人学习分享使用,禁止用于任何广告和商用目的。
6、文档遇到问题,请及时私信或留言给本站上传会员【快乐****生活】,需本站解决可联系【 微信客服】、【 QQ客服】,若有其他问题请点击或扫码反馈【 服务填表】;文档侵犯商业秘密、侵犯著作权、侵犯人身权等,请点击“【 版权申诉】”(推荐),意见反馈和侵权处理邮箱:1219186828@qq.com;也可以拔打客服电话:4008-655-100;投诉/维权电话:4009-655-100。